Wall Wash Kit Super

Placeholder

Wall Wash Kit Super

The RXSOL Wall Wash Test Kit is a complete test kit for performing a hydrocarbon test, PTT time test, chloride test, pH test, a acid wash color test and stainless steel passivation test.

Compare
SKU: RXSOL-63-6105-001 Category:

The RXSOL Wall Wash Test Kit is a complete test kit for performing a hydrocarbon test, PTT time test, chloride test, pH test, a acid wash color test and stainless steel passivation test.

The test kit consists of the following:

Lab Grade Methanol

:

10 Ltrs (2.5 ltr glass bottles X 4 Nos)

Lab Grade D.I. Water

:

10 Ltrs (5 ltrs X 2 Nos)

Pottasium permanganate

:

10 capsules

Nitric Acid 20% Soln

:

(500 ml)

Sulphuric Acid Concentrate

:

(500 ml)

Hydrochlorid Acid Concentrate

:

(500 ml)

Silver Nitrate

:

5% Solution (30 ml X 4 Btl)

Chloride Standard Solution 10 ppm

:

500 ml (1 btl)

pH Strips

:

range 0-14 (100 pcs)

Cotton Wool

:

1 bundle

Filter Paper– folded

:

100 pcs

(Whatman)Latex Gloves

:

(100 pcs)

Shoe Covers

:

(100 pcs)

Squirt Bottles– 500 ml

:

2 pcs

Thermometer

:

Range -10 to +110 deg C

Penlight with batteries

:

1 pcs

Glass sample bottles 500 ml

:

1 pcs

Funnel

:

Round – 1 pcs

Funnel Flat Sided (Stainless Steel)

:

1 Pcs

Pipette

:

plastic 3ml (10 pcs)

Nessler Tube

:

100 Ml (2 Pcs)

Measuring Cylinders 50 ml

:

(2 pc)

Nessler Tubes 50 ml

:

(2 pcs)

Volumetric Flask 500 ml

:

1 pcs

Pipette

:

glass 1ml

Pipette

:

glass 5ml

Pipette

:

glass 10ml

The details are given completely free of undertaking. Since the application lies outside our control we can not accept any liability for the results.

 

Technical Bulletin

 

PERMANGANATE TIME TEST

Source: ASTM D 1363

Scope:

This method serves as a means of detecting in alcohols or ketones the presence of impurities that reduce potassium permanganate.

Applicable to methanol, ethanol, propanol , butanol, acetone, methyl ethyl ketone andmethyl isobutyl ketone.

Summary of method:

Substances reacting with potassium permanganate in neutral solutions reduce it to manganese dioxyde, that colours the solution yellow. In the permanganate test the time required for the colour of test solution to change to that of a standard solution, is measured. The colour of the test solution changes from pink-orange to yellow-orange.

Apparatus:

Cylinders,
glass-stoppered, 50 ml;
Pipette, capable of delivering 2 ml of solution;
Clock or stopwatch.

Reagents:

Potassium permanganate solution, 0,100 g of KMnO4 per 0,5 ltr water
Uranyl nitrate – cobaltous chloride, standard solution.

This standard solution represents the colour of the terminal point to which the sample solution fades in the KMnO4 test. This solution is stable and should be kept in a 50 ml glass-stopperd cylinder, exactly the same as those in which the test is run.

Procedure:

Dissolve 0,100 g of sodium permanganate in distilled water in a volumetric flask of 500ml and fill up to the mark.  Clean a glass- stoppered 50 ml cylinder by 10 rinsings with tap water, 5 rinsings with distilled water and 3 rinsings with sample.  Fill the cylinder with sample up to the 50 ml mark and put it into a constant temperature bath, maintained at 15oC. for methanol or at 25o C. for acetone. When the sample has reached the bath temperature (about 5 minutes), add 2 ml of potassium permanganate solution, using the 2 ml pipette. Stopper the tube, invert once to mix the contents and return it to the bath.
Determine the time from addition of the potassium permanganate till the color matches that of the standard solution. Protect the tube from light during this time.
When you have finished the test, clean the sample cylinder twice with tap water and fill it with concentrated hydrochloric acid.

 

 

 

 

 

ACID WASH COLOUR OF AROMATIC HYDROCARBONS

Source: ASTM D 848

Scope:

This method covers the determination of the acid wash colour of benzene, toluene, xylene, refined solvent naphta and similar aromatic hydrocarbons.

Definition:

Acid wash colour: The colour developed in the separated acid when a sample is agitatedwith sulphuric acid under the conditions described in this method.

Apparatus:

Acid wash colour standards, numbered 0 to 14;
Shaking cylinders, 50 ml.

Procedure:

Fill a dry and clean 30 ml stoppered cylinder up to the 7 ml mark with sulphuric acid of the strength specified in Table 1 for the type of sample to be tested. Add sufficient sample to bring the total volume to the 28 ml mark. Insert the stopper, hold a finger over the stopper and give vigorous shakes with a stroke of 10 to 25 cm, shaking for a total of 150 cycles over a period of 40 to 50 seconds, that is ata rate of 3 to 3,75 cycles per second.

Allow the cylinder to stand, protected from direct sunlight, for the period of time shown in Table 1.Without further delay invert the cylinder gently once or twice to obtain a uniform colour in the acid layer and compare the colour of the acid layer with those of the standards. Make the comparison against a white background or at daylight.

Designate the colour of the acid layer by the number of the nearest matching standard and add to the number a plus or minus sign if the sample is respectively darker or lighter than the standard colour.

Note: Concentrated sulphuric acid will cause severe burns on contact with the skin. When spill, remove with plenty of water. Remark: It is advisable, when performing the test, to use clean plastic disposable gloves. TABLE 1

 

Acid strengths and standing times

Sample

 

Acid strengths

Standing time

Group 1

Benzene, all ASTM grades

Toluene, all ASTM grades

Xylene, nitration grades

Xylene 5 Degree

Xylene 10 Degree

 

 

96%

 

 

 

15 Minutes

Group 2

Xylene, industrial grade

Refined solvent Naphta

96%

5 Minutes

Group 3

Hi-flash solvent

Heavy solvent Naphta

78%

5 Minutes

 

 

 

 

 

 

 

CHLORIDE TEST
 

Principal: The principal of the test is that chloride together with a silver nitrate solution gives a milky cloudy solution.

Apparatus:

Silver nitrate solution 5%,
Chloride standard solution 1 mg chloride/ml;
Distilled water;
1  Pipette 5 ml;
1 Pipette 10 ml;
2 Funnels;
3 Nessler tubes 100 ml ;
Cotton wool ;
Filter-paper.

Procedure:

1 m2 of the surface of a tank is washed with cotton wool drenched in distilled water.

The distilled water in the cotton wool is transferred by squeezing out into a Nessler tube. The collected water is filtered off, using a funnel with filtering-paper which is placed on top of another Nessler tube.

There after the tube is filled up with distilled water to the 100 ml mark and 5 drops of a silver nitrate solution are added. The contents are mixed thoroughly. This is the test tube.

Fill another Nessler tube with 1 ml of standard chloride solution, 99 ml of water and 5drops of the silver nitrate solution. The contents are mixed thoroughly. This is the standard tube.

The two solutions are compared.

The chloride concentration in the standard Nessler tube is 10 ppm. When the turbidity of the test tube is less than the turbidity of the standard tube,   content on 1 m2 surface of the tank is less than 1 mg.

When the turbidity of the test solution in the test tube is more intensive than the  turbidity in the standard tube, the chloride content on 1 m2 surface of the tank is more than1 mg. When the turbidity in both the tubes are more or less same, the chloride content on 1 m2 surface of the tank is 1 mg. Using the total surface and the cubic content of the tank, the chloride content which can be assimilated by methanol from the surface of the tank can be calculated.

Remark: It is advisable, when performing the test, to use clean plastic disposable gloves.

 

 

 

 

 

 

HYDROCARBON TEST FOR ALCOHOLS

Scope:

This method serves a means of detecting in alcohols the presence of hydrocarbons.As the surface of a tank or container used for shipping or storage of alcohols must be freeof hydrocarbons this field-test was developed.

Summary of method:

The alcohol is mixed with distilled water. As hydrocarbons mixed with water give a milkycloudy solution, the solution of alcohol and distilled water will turn cloudy whenhydrocarbons are present.

Apparatus:

3 Nessler tubes;
2 Measuring cylinders 50 ml;
Methanol, pure;
Distilled water.

Procedure:

1 m2 of the surface of the tank is washed with cotton wool drenched in pure methanol. After each washing the methanol is squeezed out into a Nessler tube until 15 ml of methanol is collected.
 

45 ml of distilled water is added to the methanol in the tube.
The contents are mixed thoroughly.
The solution must stands for 20 minutes after shaking.

Another Nessler tube is filled with 60 ml of distilled water. This is called the ’blank’.
 

When after the standing time the solution of methanol and distilled water is not as clear as the blank, or is cloudy, it showes that the surface of the tank is not free from hydrocarbons.

Remark: It is advisable, when performing the test, to use clean plastic disposable gloves

Product Label

The RXSOL Wall Wash Test Kit is a complete test kit for performing a hydrocarbon test, PTT time test, chloride test, pH test, a acid wash color test and stainless steel passivation test.

The test kit consists of the following:

Lab Grade Methanol: ……………………………………………………. 10 Ltrs (2.5 ltr glass bottles X 4 Nos)

Lab Grade D.I. Water: …………………………………………………… 10 Ltrs (5 ltrs X 2 Nos)

Pottasium permanganate : ……………………………………………… 10 capsules

Nitric Acid 20% Soln : …………………………………………………… (500 ml)

Sulphuric Acid Concentrate : …………………………………………… (500 ml)

Hydrochlorid Acid Concentrate :………………………………………… (500 ml)

Silver Nitrate :……………………………………………………………….. 5% Solution (30 ml X 4 Btl)

Chloride Standard Solution 10 ppm : ………………………………….. 500 ml (1 btl)

pH Strips : …………………………………………………………………… range 0-14 (100 pcs)

Cotton Wool : ………………………………………………………………. 1 bundle

Filter Paper– folded : ……………………………………………………… 100 pcs

(Whatman)Latex Gloves :………………………………………………… (100 pcs)

Shoe Covers : ……………………………………………………………… (100 pcs)

Squirt Bottles– 500 ml : …………………………………………………. 2 pcs

Thermometer : …………………………………………………………….. Range -10 to +110 deg C

Penlight with batteries : ………………………………………………… 1 pcs

Wide neck glass sample bottles 500 ml : ………………………….. 1 pcs

Funnel : Round – …………………………………………………………. 1 pcs

Funnel Flat Sided (Stainless Steel) : …………………………………. 1 Pcs

Pipette : ……………………………………………………………………… plastic 3ml (10 pcs)

Nessler Tube : ……………………………………………………………… 100 Ml (2 Pcs)

Measuring Cylinders 50 ml : ……………………………………………… (2 pc)

Nessler Tubes 50 ml : …………………………………………………….. (2 pcs)

Wide neck glass sample bottles 500 ml : ……………………………… 1 pcs

Pipette : ……………………………………………………………………….. glass 1ml

Pipette : ……………………………………………………………………….. glass 5ml

Pipette : ………………………………………………………………………. glass 10ml

The details are given completely free of undertaking. Since the application lies outside our control we can not accept any liability for the results.

 

How Can we Help You
Send via WhatsApp

Enter your keyword